Symbols
Symbols
Function / Meaning
Recommended Action
Appears during measurement when
the cuff is attached loosely
Detected rate of IHB in memory
User
Previous measurements stored in memory.
Average data
Full Battery
The battery power indicator during
measurement.
Low Battery
The battery is low when it blinks.
Unstable blood pressure due to
movement during measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
PUL. DISPLAY ERROR
The pulse was not detected correctly.
Blood pressure monitor internal error
SYS.
Systolic blood pressure in mmHg
DIA.
Diastolic blood pressure in mmHg
PUL./min. Pulse per minute
AM
Data taken between 4:00 and 9:59
PM
Data taken between 18:00 and 1:59
Remove the batteries and
press the button,
and then install the batteries
again. If the error still appears
contact the manufacturer.
START
Apply the cuff correctly and
take another measurement.
Take another measurement.
Remain very still during
measurement.
Replace all batteries with new
ones when the mark blinks
The reading may yield an
incorrect value. Apply the
cuff correctly and take
another measurement.