ENGLISH 3
Symbols
Function / Meaning
Treatment
Turns the device on and off.
___________
Guide to install battery
___________
Direct current
___________
SN
Serial number
___________
Date of manufacture
___________
Type BF: Device, cuff and tubing
are designed to provide special
protection against electrical
shocks.
___________
Appears when measurement is in
progress and flashes when the
pulse is detected.
Measurement is in progress -
- remain still.
Appears when the battery voltage
is too low for the device to work
properly.
Replace all batteries with
new ones.
Appears if the systolic and diastolic
measurements are within 10
mmHg of each other.
Verify that the air hose is
properly connected and that
the exhaust rate is between 2
and 5 mmHg.
Check for air leaks.
Appears if the pressure value is
unstable due to movement during
the measurement.
Try the measurement again.
Remain very still during the
measurement.
Appears if the pressure value
does not increase during inflation.
Verify that the air hose is
properly connected and that
the exhaust rate is between 2
and 5 mmHg.
Check for air leaks.
Pulse beats per minute
___________
The medical device label by the
EC directive
___________
Registered with the Australian
Communication Authority by the
Trademarks office.
___________
Summary of Contents for UA-767