AND UA-1200BLE Instruction Manual Download Page 1

Digital Blood Pressure Monitor 

Model UA-1200

BLE

 

 

Instruction Manual 

Original 

Manuel d

instructions 

Traduction

Manual de Instrucciones 

Traducción 

Manuale di Istruzioni 

Traduzione 

使用手冊

 

翻譯

 

 

ﺕﺍﺩﺎﺷﺭﻹﺍ ﻝﻳﻟﺩ

 

ﺔﻣﺟﺭﺗﻟﺍ

 

1WMPD4003740

 

English 

Français 

Español 

Italiano 

中文 

ﻟﺍ

ﻳﺑﺭﻌ

Summary of Contents for UA-1200BLE

Page 1: ...odel UA 1200BLE Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione 使用手冊 翻譯 ﺍﻹﺭﺷﺎﺩﺍﺕ ﺩﻟﻳﻝ ﺍﻟﺗﺭﺟﻣﺔ 1WMPD4003740 English Français Español Italiano 中文 ﺍﻟ ﻌﺭﺑﻳ ﺔ ...

Page 2: ...Identification 6 Contents 6 Symbols 7 Using the Monitor 10 Measurements 16 What is an The IHB AFib Indicator 20 What Is The AFib 20 WHO Classification Indicator 20 About Blood Pressure 21 Settings by App 23 Setting by Monitor 24 Troubleshooting 25 Maintenance 26 Technical Data 26 ...

Page 3: ... their respective owners Preliminary Remarks This device conforms to the European Directive 93 42 EEC for Medical Products This is made evident by the mark of conformity 0123 The reference number to the involved notified body Hereby A D Company Limited declares that the radio equipment type UA 1200BLE is in compliance with Directive 2014 53 EU The full text of the EU declaration is available at th...

Page 4: ...nd batteries are not treated as ordinary household waste and must be disposed of according to the applicable local regulations Do not modify the device It may cause accidents or damage to the device To measure blood pressure the wrist must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery This may cause pain numbness or a temporary red mark to the wrist This con...

Page 5: ... mistake by infants Do not touch the batteries and the patient at the same time That may result in electrical shock Wireless communication devices such as home networking devices mobile phones cordless phones and their base stations walkie talkies can affect this blood pressure monitor Therefore a minimum distance of 1 meter 3 2 feet should be kept from such devices Unplug the ac adapter when not ...

Page 6: ...nose the measurement results and start treatment by yourself Always consult your doctor for evaluation of the results and treatment Do not apply the cuff on wrist with an unhealed wound Do not apply the cuff on wrist receiving an intravenous drip or blood transfusion It may cause injury Do not use the device where flammable gases such as anesthetic gases are present It may cause an explosion Do no...

Page 7: ...English 6 Parts Identification Contents 1 Blood Pressure Monitor 1 Instruction Manual 1 Charging Cable OLED RED LED Start Switch Cuff Reset Switch USB Connector ...

Page 8: ...protection provided by enclosures in accordance with IEC 60529 This device is protected against solid foreign objects of 12mm diameter and greater such as a finger This device is not protected against water WEEE label SN Serial number BT Bluetooth address Refer to instruction manual booklet To indicate generally elevated potentially hazardous levels of non ionizing radiation or to indicate equipme...

Page 9: ...not detected correctly Apply the cuff correctly and take another measurement Err 1 Unstable blood pressure due to movement during measurement The systolic and diastolic values are within 10mmHg of each other Take another measurement Remain still during measurement Err 3 The pressure value did not increase during inflation The cuff is not applied correctly Apply the cuff correctly and take another ...

Page 10: ... reading may yield an incorrect value Take another measurement Remain still during measurement Appears during measurement when the cuff is attached loosely The reading may yield an incorrect value Apply the cuff correctly and take another measurement Firmware is updating Appears when irregular heart beat is detected ...

Page 11: ...d to the app on next connection 3 Multiuser Mode Measurement In app Up to 6 users 5 accounts and 1 visitor can use and store data on the app Each user must have a mobile device paired with the monitor 4 Guest Mode Measurement In app User can take a measurement as Guest without account registration in the A D Connect app The data is saved on the mobile device memory but not in your account 5 Visito...

Page 12: ... be sure to charge the device completely upon first use after purchase 1 Remove the USB connector cover 2 Connect the charging cable 3 During charging the red LED will illuminate 4 When charging is complete the red LED will turn off 5 Once charging is complete remove the charging cable ...

Page 13: ... inadequately trained personnel could result in a hazard such as a fire or explosion Wireless Function In the unlikely event that this monitor causes radio wave interference to a different wireless station change the location where this monitor is used or stop use immediately Be sure to use in a location where visibility between the two devices that you want to connect is good The connection dista...

Page 14: ...r any damages incurred due to impaired operation or data loss etc that occur through the use of this product This product is not guaranteed to connect to all Bluetooth compatible devices Bluetooth Transmission This product is equipped with a Bluetooth 4 1 wireless function and can connect to the following Bluetooth devices Apple iOS v 8 0 or newer Android OS 4 4 4 5 1 6 0 and 7 0 iPhone 4s or newe...

Page 15: ...can be added anytime The user who has the lowest frequency of measurement times will be cleared once the device has reached the 5 user max Note If new user uses the same mobile device then re pairing is not required as this has been complete Note Default is in visitor non memory mode with out pairing CAUTION The communication distance between this monitor and the receiver device is about 33 feet 1...

Page 16: ...ood pressure measurement Sit comfortably on a chair Rest your arm on the table Do not cross your legs Keep your feet on the floor and straighten your back Relax for about five to ten minutes before measurement Place the center of the cuff at the same height as your heart Remain still and keep quiet during measurement Do not measure immediately after physical exercise or a bath Rest for twenty or t...

Page 17: ...the pressure value once inflation starts 6 When the measurement is complete the app and the monitor will display results similar to the icon on the right Systolic diastolic and pulse rate will be displayed in rotation Note Press the start button again to turn the monitor off The monitor is also provided with an automatic power shut off function Allow at least one minute between measurements on the...

Page 18: ...er 3 seconds 4 Zero 0 is displayed blinking briefly Then the display changes as measurement begins 5 The cuff starts to inflate It is normal for the cuff to feel tight Monitor displays the pressure value once inflation starts 6 When the measurement is complete the monitor will display result similar to the icon on the right Systolic diastolic and pulse rate will be displayed in rotation Note Press...

Page 19: ...urement 1 Up to 5 people can use and store readings on the app Each user must have a mobile device paired with the monitor See pg 14 for pairing instructions 2 If you are taking the reading without the app press the Start button on the monitor to switch between usernames Select the desired user before each reading Data will transfer to the app on your next connection Note Usernames will be display...

Page 20: ...stress the measurement will reflect this stress as a higher or lower than normal blood pressure reading and the pulse reading will usually be faster than normal An individual s blood pressure varies constantly depending on what you are doing and what you have eaten What you drink can have a very strong and rapid effect on your blood pressure This device bases its measurements on the heartbeat If y...

Page 21: ...ting your physician if you see this IHB AFib indicator frequently What Is The AFib The heart contracts due to electrical signals occurring in heart and sends blood through the body Atrial fibrillation AFib occurs when the electrical signal in the atrium becomes confused and leads to disturbances in the pulse interval AFib can cause blood to stagnate in the heat which can easily create clots of blo...

Page 22: ...ressure if left unattended can cause many health problems including stroke and heart attack Hypertension can be controlled by altering lifestyle avoiding stress and with medication under a doctor s supervision To prevent hypertension or keep it under control Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups Maintain proper weight Why Measure Blood Pressure a...

Page 23: ... Organization WHO as shown in the chart Blood Pressure Variations An individual s blood pressure varies greatly on a daily and seasonal basis It may vary by 30 to 50 mmHg due to various conditions during the day In hypertensive individuals variations are even more pronounced Normally the blood pressure rises while at work or play and falls to its lowest levels during sleep So do not be overly conc...

Page 24: ...erval is useful Default for single user is set at 1 second and for multiple user is set at 5 seconds Remove User Allows user to remove himself as a user in the BP monitor Check Software Update Enables users to update the monitor with the most recent Bluetooth software Unpair Allows user to unpair his mobile device with the BP monitor This can also be accomplished via the Bluetooth setting in the m...

Page 25: ...ency Measurement can be done in airplane mode and data will transfer to the app on next connection 1 Hold the Start button and the airplane mode select screen will display Tap the Start button to switch between ON and OFF 2 Select ON and wait 3 seconds to enter airplane mode In airplane mode the airplane icon to the left will be illuminated ...

Page 26: ... heartbeat the device may have difficult in determining your blood pressure The cuff position is not correct Sit comfortably and still Place your arm on a table with your palm facing upward and the wrist cuff at the same level as your heart Other The value is different from that measured at a clinic or doctor s office At a clinic or doctor s office apprehension may cause an elevated reading Home m...

Page 27: ...ice inspected every 2 years to ensure proper functioning and accuracy Please contact the authorized dealer in your area or A D for maintenance Technical Data Type UA 1200BLE Measurement method Oscillometric measurement Measurement range Pressure 0 299 mmHg Systolic pressure 60 279 mmHg Diastolic pressure 40 200 mmHg Pulse 40 180 beats min Measurement accuracy Pressure 3 mmHg Pulse 5 Arm circumfere...

Page 28: ...ions 10 C to 40 C 15 RH to 85 RH 800 hPa to 1060 hPa Transport Storage conditions 20 C to 60 C 10 RH to 95 RH 700 hPa to 1060 hPa Dimensions Approx 44 W x 126 H x 26 D mm Weight Approx 205 g Ingress protection Device IP20 Applied part Cuff Type BF Useful life 5 years when used six times a day Included items UA 1200BLE 1PC Instruction Manual 1PC Charging cable KSUSB 2C 1PC Option AC adapter AX TB26...

Page 29: ...English 28 MEMO ...

Page 30: ...9 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 OOO A D RUS OOO ЭЙ энд ДИ РУС 121357 Российская Федерация г Москва ул Верейская дом 17 Business Center Vereyskaya Plaza 2 121357 Russian Federation Moscow Vereyskaya Street 17 тел 7 495 937 33 44 факс 7 495 937 55 66 A D Technology Trading Shanghai Co Ltd 爱安德技研贸易 上海 有限公司 中国 ...