11
E
L
N
E
GND
N
L
GND/PE Earth
Ground
Yellow-Green mixed wire is used for
grounding! Not connected to zero wire!
Please connect according to picture or other
correct way. Please disconnect mains
power when connecting!
Model
CUT40i
Power supply
1- phase AC 220V±15%, 50/60Hz
Min. power capacity (KVA)
9
Input volt.
protection
(A)
Fuse
50
Circuit breaker
60
Min. cable size
(mm
2
)
Input cable
2.5
Output cable
16
Protective GND wire
2.5
Table 4-5-1: Input power supply cable installation
The connection between input cable and distribution box (Fig. 4-5-2)
Warning!
- Never connection when equipment is power on!
- The connection must be carried out by a qualified electrician!
- Do not connect two units of power sources to the same circuit breaker!
- Connected to the correct input voltage, circuit breaker, input cable as per the
specification on Table 4-5-1.
Fig. 4-5-2: Connection between input power supply cable and switch box