TOC Validation
The principles and procedure for performing a TOC validation are essentially the same as
those used to perform a calibration. The intent of validation is to check the accuracy of the
calibration, thereby ensuring that the Anatel A643a measures samples reliably. Validation can
be either a single-point analysis or multipoint analysis and the standard concentrations are
user-selected.
Note:
Refer to Standard Operating Procedure (SOP) #643–5/5A for complete instructions on
performing this procedure.
To perform a TOC validation:
1)
With the desired Analyzer selected in any View, press the
Manual Key
to access its
menu selections.
2)
Use the
Up
and
Down Keys
to specify
Calibrations
.
3)
Press
Enter
to display its submenu.
CAUTION:
In a networked environment (see
“Network Installation” on page 39
), it is important to confirm
the instrument’s Serial Number before proceeding. Operation of the wrong Analyzer could
result in water leakage from the sample vessel when it is removed during this procedure.
4)
Use the
Up
and
Down Keys
to specify
Validate
. If the chosen Analyzer is connected to
an A-Net network on which other instruments reside, the analyzer will prompt to confirm
that it is the correct instrument by verifying its serial number. The Analyzer will begin a
temperature stabilization process, its duration depending on the sample water
temperature. The actual time required is displayed and counted down.
5)
Press
Enter
to continue and wait while the sample line is drained.
6)
Specify the validation standard levels to be performed. A 500 ppb TOC sucrose
concentration level is enabled by default, indicated by the checkmark preceding it.
Enable any other levels by pressing the
Up
and
Down Keys
to select the desired
concentration, then pressing
Enter
to edit and
Up
and
Down
toggle the checkmark on
and off. At least one validation level must be selected.
Fig 6-9 :
TOC Validation
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......