If TOC analysis is complete before the set
Cycle Time
has elapsed, the instrument goes into
an Idle State and awaits the start of the next analysis. During this
Idle Time
, Sample Valve 1
either allows or prevents sample water flow, depending on the Analyzer’s
Sample Mode
setting (see
). When the setting is
Continuous
, the measurement
cell is continuously purged until the conclusion of the
Cycle Time
; when
Water Saver
is
selected, sample flow through the Analyzer is suspended until initialization of the subsequent
Sample Time
. If the
Cycle Time
is less than the duration of the analysis, the next analysis
begins immediately. Thus, in order to perform continuous analyses, the Analyzer's
Cycle Time
should be set to the default of
0
minutes. This setting is recommended unless a fixed and
lengthy (hourly) analysis is required.
5.5.1
Auto TOC Software Setup
There are several parameters which define Anatel A643a operation in the Auto TOC Mode.
These variables include setting the
Sample
and
Cycle Times
, as well as the instrument’s
Sample Mode setting. These time parameters are accessed through the instrument’s
Times
& Cycle Menu
and determine the duration of its analysis. As noted, the analysis
Cycle Time
is comprised of
Sample
,
Oxidation
and, in some instances, an
Idle Time
.
To access the Anatel A643a’s
Times & Cycle Menu
:
1)
With the desired Analyzer selected in any View, press the
Setup
Key
to display its
submenu.
2)
Use the
Up
and
Down Keys
to specify
Sensor Setup
.
3)
Press
Enter
to access it submenu.
4)
Specify
Analysis Setup
from among the available selections and press
Enter
to
display its options.
The
Times & Cycle Menu
offers the following instrument-specific analysis parameters:
•
Sample Tim
e is the period of time water flows through the Analyzer before its internal
sample valve closes and sample analysis begins.
•
Cycle Time
is the minimum length of time from the beginning of one analysis to the
beginning of the subsequent analysis.
Fig 5-20 :
Auto TOC Software Setup
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......