2.2
Pre-Installation Procedures
Preliminary setup procedures should be conducted on the Anatel A643a components prior to
installing them on the water system to help assure their proper operation. Verification also is
easier at this stage than when the components have been mounted and cabling has been run.
Suggested pre-installation procedures are:
•
Unpacking, visually inspecting the equipment and verifying the packing list.
•
Powering up the Analyzer to verify that all the components are operational (refer to
“Specifications and Accessories” on page 191
for the specific power requirements of
each component).
•
Setting up the operational software by entering the System Time and naming the
Analyzer.
Note:
The terms “
Anatel A643a
” and “Analyzer” refer generally to all A643a TOC Analyzer models.
2.2.1
Unpacking & Inspection
Upon receiving the Anatel A643a Analyzer, inspect the shipping container(s) for any signs of
external damage before unpacking the components. Carefully remove the equipment, verify
the enclosed packing list and check once again for obvious damage.
Each Anatel A643a Installation Kit contains:
•
Power Cord
•
Network “Tee” & Terminator
•
15 Micron Inlet Filter
• 5’ 1/4” OD Sample Tubing
•
10’ 1/8” OD Drain Tubing
•
1/4” Tube x 1/4” MNPT Fitting
•
Operator Manual and Standard Operating Procedures (SOPs)
If damage is apparent, notify the freight carrier immediately: Claims must be filed by the
customer. Also notify Hach Ultra or your distributor of any such problems. Include the Serial
Number of the damaged unit(s) and your purchase order number with all factory
communications.
2.2.2
Operational Verification
The next pre-installation step is to establish communication connections between the C80
Controller and its associated Anatel A643a–S Analyzer (these connections are internal to the
Anatel A643a–P).
Each Anatel A643a–S Analyzer and C80 Controller is supplied with a 3-foot local cable, a twin
BNC “tee” connector, and a passive terminator. The C80 Controller also includes a 10-foot
trunk cable to facilitate communications connections.
Note:
Additional twin-axial trunk cable, pre-wired with BNC plugs at each end, is available from Hach
Ultra in 10-, 25-, 50-, 100-foot and custom lengths.
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......