Anatel A643a - Serial Communications
151 of 220
RP
S -
J
u
ne
20
07
- E
d
it
io
n 1
2
Anatel
Operator Manual
12.3.3 Data Read Commands
The Data Read functions return multiple values, the string terminated with a <cr><lf>.
Note:
If the Anatel A643a Analyzer is in an operational mode that does not generate a TOC result
(modes 6 through 12), the TOC, TOC Alarm Percentage, TOC Trend, Profile (Curve) Type and
Oxidation Time are not returned.
Note:
Data normally is returned only in response to issuance of the RD or RE commands. When
collecting data with a serial communications program such as Windows 95 Hyperterminal,
however, it is more convenient to have the data displayed automatically. This can be done
using the SA command, which then displays data whenever they are reported.
20mA Output On Alarm (0 = minimum value,
1 = unchanged, 2 = maximum value)
9.2.2.4
TOC Mode Paper Saver Percentage Change (1–99)
5.5.2
Purge Mode Paper Saver Percentage Change (1–99)
5.6.1
Purge Mode Interval Between Printouts (hh:mm:ss)
5.6.1
Zero-Scale TOC Range (ppb)
9.2.2.3
Full-Scale TOC Range (ppb)
9.2.2.3
User TOC Calibration Slope
(Read-Only)
External DAC Zero-Scale Conductivity/Resistivity Range
(Units depend on
Conductivity/Resistivity Units of Measure)
9.3.2.3
External DAC Full-Scale Conductivity/Resistivity Range
(Units depend on
Conductivity/Resistivity Units of Measure)
9.3.2.3
External DAC Zero-Scale Temperature Range (°C)
9.3.2.4
External DAC Full-Scale Temperature Range (°C)
9.3.2.4
Digital Control (0 = disabled, 1 = enabled)
10.2
Log (0 = Data & Audit, 1 = Audit Only)
8.3.1
User Conductivity Calibration Slope
(Read-Only)
Example:
"HR 00:01:00 00:00:00 500 0 0 1
SENSOR_NAME 0 0 0 0 0 0 0 0 0 0 1 1 00:01:00
0 1000 1.0 0 20 5 95 0 1.0 <cr>"
SY
Set Time (MM:DD:YY:hh:mm:ss format)
2.2.3.1
Example:
"SY 03 04 1993 12 33 00 <cr>"
Table 12-2 :
Parameter Set Commands (Continued)
Command Function
Section
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......