12.3
Anatel A643a Command Set
Serial interface commands consist of Mode Set, Parameter Set, Data Read, Data Logger and
Data History functions. The commands consist of twocharacter ASCII text mnemonics. Some
commands also require one or more arguments, each delimited by at least one space (ASCII
32; 20 Hex), followed by the command mnemonic. Each command string is terminated with a
carriage return (ASCII 13; 0D Hex).
Note:
Anatel A643a
Analyzers respond to either uppercase or lowercase commands.
The Anatel A643a Analyzer will respond with an “OK>” prompt after the command has been
accepted. In addition, Data Read and Parameter Set commands elicit a reply that consists of
one or more numeric or text values, each delimited by at least one space, and terminated by a
carriage return, linefeed pair (ASCII 13,10; 0D,0A Hex).
Both commands and replies use a “free field” format, so the number of delimiting spaces and
length of each argument or data field may vary. If a command is rejected due to invalid syntax,
for instance, refusal is indicated by a question mark (ASCII63; 3F Hex) reply preceding the
“OK>” prompt.
The notations used to represent command arguments below are as follows:
12.3.1 Mode Set Commands
The Mode Set commands are used to determine the instrument’s operational mode.
hh:mm:ss
hours:minutes:seconds
n
decimal number (e.g., “1.234)
i
integer (e.g., “1”)
s
text string (e.g., “SENSOR_NAME”)
b
binary flag (“1” or “0”)
Table 12-1 :
Mode Set Commands
Command Function
Comment
MC
Self-Clean Mode
Valve open, lamp on.
MD
Auto TOC Mode
Continuous TOC analyses.
ME
Clear Code Log
Erases Codes from memory.
MO i
Start TOC Analysis One or more TOC analyses, then goes to Idle State. The default is
one analysis cycle if an argument (“i”) is not specified.
MP
Purge Mode
Valve open, lamp off, shows resistivity/conductivity and temperature.
MZ
Idle State
Places the Analyzer in the Idle State (if present).
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...