Anatel A643a Alarms - Anatel A643a
R
PS
-
June
20
07
-
Ed
it
io
n
12
Operator Manual
Anatel
When the uncompensated limit is exceeded, an alarm is generated and several things occur:
•
The LED corresponding to the Anatel A643a Analyzer’s Channel ID flashes red on all of
its associated C80 Controllers.
•
The alarm is logged and becomes available for display on any of those Controllers.
•
If the Analyzer is attached to a local or a Gateway printer, a hardcopy of the condition is
generated containing the time of the excursion, the measured conductivity and
temperature, and the USP conductivity limit exceeded. For example:
12:07 Alarm 1.45 µS @ 33.2C >[1.4 @ 30C]
Excursions reported in subsequent analyses are output only at the normal user-specified
interval (or in response to a percent change in the readings) unless the sample water
conductivity falls below the specified limit. In that case, a new alarm report is output to
ensure that each limit excursion is recorded.
•
The Sensor’s
Digital Output #2
is asserted to allow alarm annunciation to an external
device (see
7.4
Alarm Acknowledgement
A limit excursion is acknowledged by specifying the errant Analyzer and then pressing the
Alarm Key
. Acknowledgement stops the Controller’s LCD from flashing, stops the flashing red
Channel ID, and silences the audible warning. If more than one alarm has been detected, all
must be acknowledged before these fault indications are terminated. The red Channel LED is
restored to green when the Analyzer’s TOC level falls below its set limit.
Acknowledgement presents the following alarm information on the Controller’s LCD display:
Fig 7-16 :
Uncompensated Conductivity Alarm
Line 1
The affected instrument’s Channel ID and its userentered
Name
.
Line 2
The cause of the alarm.
Line 3
The date and time the alarm was detected.
Line 4
The current number of alarms logged for the displayed Analyzer.
Summary of Contents for A643
Page 1: ...Operator Manual ANATEL A643a TOC ANALYZER...
Page 2: ......
Page 18: ...18 of 220 Introduction Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 80: ...80 of 220 Anatel A643a Setup Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 110: ...110 of 220 Anatel A643a Alarms Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 120: ...120 of 220 Printer Output Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 138: ...138 of 220 Analog Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 144: ...144 of 220 Digital Inputs Outputs Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 154: ...154 of 220 Serial Communications Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 188: ...188 of 220 Troubleshooting Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 190: ...190 of 220 Service Procedures Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 212: ...212 of 220 Glossary Anatel A643a RPS June 2007 Edition 12 Operator Manual Anatel...
Page 219: ...Anatel A643a User Notes 219 of 220 RPS June 2007 Edition 12 Anatel Operator Manual User Notes...
Page 221: ......
Page 222: ......