background image

Basic information 

Biometra TOne 

Version 01.21 

1

 

Basic information 

1.1

 

Notes on this Setup Instruction 

As an excerpt from the detailed device instructions, these setup instructions contain the 
first steps of installation and the safety instructions to be observed. For all further steps, 
please apply the detailed operating instruction, which is supplied in electronic form on a 
USB stick. Alternatively, you can also find the operating instruction on the Analytik Jena 
homepage in the download area of the corresponding device. 

Biometra TOne is intended for operation by qualified specialist personnel observing the 
device operating manual for the Biometra TOne software. 

The operating manual contains information about the design and operation of Biometra 
TOne and provides personnel familiar with PCR technology with the necessary know-
how for the safe handling of the device and its components. The operating manual fur-
ther includes notes on the maintenance and servicing of the device and potential causes 
and remedies of any faults. 

This instruction manual applies to all Biometra TOne devices from software version ME 
1.70 

 RE 1.46. 

 

Safety instructions

 are indicated by pictographs and signal words. The type and source 

of the hazard are stated together with notes on preventing the hazard. 

 

The operating manual uses the following symbols and signal words to indicate hazards 
or instructions. Safety instructions are always placed before an action. 

 

WARNING 

Indicates a potentially hazardous situation which might cause fatal or very serious inju-
ries (deformities). 

 

CAUTION 

Indicates a potentially hazardous situation which might cause light or minor injuries. 

 

NOTE 

Provides information on potential material or environmental damage. 

1.2

 

Intended use 

The Biometra TOne is an end-point thermal cycler. It is designed to amplify nucleic acids 
by repeated cycles of heating and cooling using DNA polymerases in a PCR reaction.  

The thermal cycler is developed for Research Use Only (RUO). 

The TOne combines modern design with user-friendly software. The user interface con-
sists of a touchscreen with a graphical or spreadsheet display that shows the time, the 
status and the temperature program for each run. Using the touchscreen keyboard, 

Conventions 

Symbols and signal words 

Summary of Contents for Biometra TOne

Page 1: ...Setup Instructions Biometra TOne High performance thermal cycler for DNA amplification by PCR...

Page 2: ...a com These setup instructions are valid from software version ME 1 70 RE 1 46 For a proper and safe use of this product follow the instructions Keep the setup instruction for future reference Microso...

Page 3: ...eration 7 2 6 1 Summary of safety instructions 7 2 6 2 Safety instructions protection against explosion and fire 8 2 6 3 Safety instructions electrical equipment 8 2 6 4 Handling of samples auxiliary...

Page 4: ...lts This instruction manual applies to all Biometra TOne devices from software version ME 1 70 RE 1 46 Safety instructions are indicated by pictographs and signal words The type and source of the haza...

Page 5: ...horten program run times Deviations from the intended use described in this short instruction and in the operating manual result in limitations of warranty and liability during a damage event Damage t...

Page 6: ...to the inside of the heated lid and the rear of the device Symbol Meaning Notes Caution Hazard point This symbol is located on the inside of the heated lid and on the rear of the device Proceed with d...

Page 7: ...s The operating manual must be accessible to the operating and service personnel at all times 2 5 Safety instructions transport and installation The following has to be observed Drain the sample block...

Page 8: ...r service 2 6 2 Safety instructions protection against explosion and fire Biometra TOne must not be operated in an explosive environment Biometra TOne must not be operated with flammable explosive or...

Page 9: ...of the examined biological material 2 7 Safety instructions maintenance and repair Biometra TOne is usually repaired by the Biometra GmbH Analytik Jena AG service de partment or by authorized and trai...

Page 10: ...n a se cure and dry environment Consider the specific ambient conditions 15 C to 30 C max 70 air humidity max 2000 m above sea Allow the Biometra TOne thermal cycler to equilibrate to room temperature...

Page 11: ...ble Check the operating voltage before switching on the device 3 4 Setting the operating voltage WARNING Danger of electric shock Before commissioning the device make sure that the voltage selector se...

Page 12: ...4 1 Biometra TOne front view Fig 1 Biometra TOne front view 1 Dial for adjusting the contact pressure 2 Lid closing button 4 Display 6 USB port 4 2 Biometra TOne rear view Fig 2 Biometra TOne PCR mod...

Page 13: ...her information please check out the operating manual of the device These manuals are included in electronic form on USB flash drive Alternatively you can find the operating manual on the Analytik Jen...

Reviews: