Fig. 70 Primer annealing temperature calculator screen
To adjust the primer or salt concentration press the
Edit concentration
button.
Edit concentration
In the next screen the primer concentration, salt concentration and magnesium concen-
tration can be adjusted. Touch the corresponding input field and enter values using the
keypad. The concentrations have the following default values and ranges:
Function
Default value
Range
Primer concentration
200 nmol/l (nM)
0-2.000 nmol/l
Salt concentration
50 mmol/l (mM)
0-200 mmol/l
Magnesium concentration
0 mmol/l (mM)
0-200 mmol/l