AS 21hp, AS 10e
Dismounting and reassembling
17
Loosen the 2 screws with an Allen key (1.5 mm) and remove the Z-drive.
The reassembling of the Z-drive will be done in reverse order.
Perform the final test according to chapter 4.
3.3
Exchange of X-drive (carrousel guide)
Move the sampler into the service position.
Fig. 19 Unplugging the electrical connector from the main board (exchange of X-drive)
Unplug the tray sensor and X-sensor from main board.