Maintenance and care
InnuPure C16
Edition 06/2015
51
8
Maintenance and care
The InnuPure C16 is mainly maintenance-free. Care and maintenance are limited to cleaning
the device, changing the fuses and installing software updates.
CAUTION!
The device cover may only be opened by the technical customer service of Analytik Jena AG
or its authorized agency. Disconnect the mains plug before opening the device cover!
8.1
Cleaning the InnuPure C16
CAUTION! Danger of electric shock!
Before cleaning the device with disinfection agents, disconnect the mains plug from the con-
nection of the InnuPure C16!
The InnuPure C16 must only be recommissioned once it is completely dry!
Clean the InnuPure C 16 as follows:
Avoid contamination by handling sample substances with care.
Wipe spilled samples or reagents immediately with an absorbent cloth or piece of
paper.
If the InnuPure C16 is used for the analysis of infectious material, great care must
be taken because the InnuPure C16 cannot be decontaminated as a whole device.
Visible contamination must be removed immediately using suitable detergents,
making sure that no solvent enters the inside of the device.
The drip tray can be removed from the device for cleaning with the door open.
Caution!
The magnet heating elements must not be in the up position.
The adapters for accepting the reagent strips can be autoclaved and are also suit-
able for dip and spray disinfection.
The sample tray is also suitable for dip and spray disinfection.
We recommend the following disinfection agents for the device interior
Descosept AF
Dr. Schuhmacher GmbH
Meliseptol HBV – cloths
B. Braun
These are only to be used for wiping disinfection. If using a spray head, first apply
the disinfection agent on suitable cloths.
We recommend the following disinfection agent for the front screen made of acrylic
glass
Descosept Spezial
Dr. Schuhmacher GmbH
This is only to be used for wiping disinfection. If using a spray head, first apply the
disinfection agent on suitable cloths.