First commissioning
InnuPure C16
Edition 06/2015
23
5.1.3
Energy supply
WARNING!
The InnuPure C16 must only be connected to a properly grounded mains outlet in accord-
ance with the voltage specifications on the type plate!
The InnuPure C16 is operated on single-phase alternating current. The installation of the
electrical equipment in the laboratory must comply with the standard DIN VDE 0100. At the
connection point an electrical current in accordance with the standard IEC 38 must be avail-
able.
5.2
Connecting the InnuPure C16 and switching it on/off
IMPORTANT
Retain the transport packaging! Return transport for service must be in the original packag-
ing. This alone prevents transport damage.
Connecting the InnuPure C16
1.
Take the InnuPure C16 out of its packaging.
2.
Check the device for completeness and soundness.
3.
Insert the control unit HID-Pro 320 into the interface on the front of the InnuPure C16 (1
in Figure 2 p. 16)
4.
Connect the mains cable at the rear.
The InnuPure C16 is now ready for operation.
Switching on the InnuPure C16
ATTENTION!
When switching on the InnuPure C16, no USB stick must be connected to the control unit
HID-Pro 320. Self-created user accounts on the HID-Pro 320 could be deleted.
1.
Switch the InnuPure C16 on from the mains switch at the equipment backplate.
2.
Activate the InnuPure C16 via the touch sensor on the front of the device. Touch the
sensor for approx. 3 s until the status indication currently flashing red/green changes to
green only and the control unit HID-Pro 320 switches on.
Note:
The switch is a capacitive button that could be damaged by strong pressure. Only touch
the button lightly.
The InnuPure C16 is initialized automatically. The process takes approx. 30 s.
The end of initialization is signaled by a beeping sound. Only then may en-
tries be made via the control unit HID-Pro 320.