Sub MkIIIF
User Manual
Int. Approved
Document ref: ASF3-800-08
November 2018
Page 19 of 79
Copyright © 2009 Analox Ltd. All Rights Reserved.
Commercial in Confidence
Figure 4
REM2 Remote Sensor Dimensions
The REM3 shares the basic dimensions of the REM1, although in standard form it does not have a
CO2 sensor fitted.
All enclosures intended for hyperbaric/hypobaric conditions are fitted with a vent to prevent
implosion/explosion. Ensure that the vent is not obstructed.
Ensure that the remote sensor units are positioned sensibly with regard to the gases being
monitored, and the protection to personnel that the sensors are to provide. Gas sensors should
not be mounted in locations where air circulation is likely to be restricted. Access for calibration
must also be provided.
REM3’s being used as part of oxygen injection systems should not be positioned too closely to the
injection point(s) in the chamber.
Note also that sensor modules with displays are usually supplied with a protective film over the
front panel label and display viewing window. This protective film is intended to be removed to
provide better visibility of the displays.