EII CO
Technical User Manual
Analox Ltd
15 Ellerbeck Court, Stokesley Business Park
North Yorkshire, TS9 5PT, UK
T: +44 (0)1642 711400 F: +44 (0)1642 713900
W: www.analox.net
E: [email protected]
Page 1: ...EII CO Technical User Manual Analox Ltd 15 Ellerbeck Court Stokesley Business Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net...
Page 2: ......
Page 3: ...Ref P0107 801 00 February 2011 Page 4 List of Contents 1 Introduction 6 1 1 Scope of this Manual 6 1 2 About the EII CO 6 2 Calibration 7 2 1 Items required 7 2 2 Calibration procedure 7 3 Safety Inf...
Page 4: ...EII CO Technical User Manual Document Ref P0107 801 00 February 2011 Page 5...
Page 5: ...esigned and equipped with several features to ensure ease of use and reliability The instrument has been designed to be held in the left hand to facilitate checking your tank It is fitted with a large...
Page 6: ...a flat blade screwdriver loosen the 4 M2 5 slotted screws located on the front cover these are captive screws so can not be fully removed from the cover The gasket seal is designed to prevent water le...
Page 7: ...e screws and black fibre washers Remove the rear cover 5 Screw the flow indicator on to the zero calibration cylinder synthetic air and fit the flow adaptor to the tubing Flow indicator fitted Flow ad...
Page 8: ...to read 000 Display set to 000 9 Close the control valve of the flow indicator Remove the control valve from the zero calibration cylinder and fit it to the span calibration cylinder 10ppm carbon mono...
Page 9: ...Place the lanyard loop over the dowel pin then fit the back cover in place making sure the CO sensor lead runs down the rear of the battery compartment as shown Affix the cover using the two M3x16mm...
Page 10: ...2011 Page 11 15 Feed the CO sensor cable so that it curls around the inside of the middle section fit the red front section Secure in place with the four M2 5 slotted screws taking care not to trap or...
Page 11: ...se Obtain medical advice if continued irritation Ingestion Corrosive May cause sore throat abdominal pain nausea and severe burns of the mouth throat and stomach and may be fatal If swallowed DO NOT I...
Page 12: ...Bump test gas kit 10ppm CO in balance air Providing around 20 bump tests from a standard bump test kit SA7CAN510 Span calibration gas cylinder 10ppm carbon monoxide in balance air SA7EIICOCALKIT Zero...
Page 13: ...P0107 801 00 February 2011 Page 14 5 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for information on the disp...