Compressor connection
Connection of the Analox CO
Clear
to the compressor will require a pressure regulator. The
pressure regulator must fit the following specifications:
•
Maximum inlet pressure 200Bar
•
Outlet pressure 0-3 BarG
For systems where 1/4" pipe work is used the push fitting in the sensor housing must be
swapped out.
Remove the sensor housing and carefully remove the 6mm push fitting and replace with
the provided 1/4" push fitting and tighten, replace the sensor housing into the bulkhead.
To ensure trouble free operation of the CO
Clear
a 2-stage pressure regulation should be
used. The output of the selected regulator should be set a 2 bar