SL 5107 Rev. A.
1
iReceptal Digital Surgical Suction System
Site Preparation, Installation, and Maintenance Guide
100
~
240 VAC Docking Station
REF iRD301-01
Instructions For Use
R
x
ONLY
Page 1: ...SL 5107 Rev A 1 iReceptal Digital Surgical Suction System Site Preparation Installation and Maintenance Guide 100 240 VAC Docking Station REF iRD301 01 Instructions For Use Rx ONLY ...
Page 2: ...Instructions 9 To Install the Docker 9 1 To Install the Docker and Connect Utilities 10 2 To Apply Power Water and Detergent 11 To Test the Docker 11 1 To Prepare the Rover 12 2 To Prepare the Dock 13 3 To Perform a Wash Cycle 13 To Shut Down the Docker 14 Inspection and Maintenance 14 Add the Detergent 15 Cleaning 16 Recommended Equipment 16 To Wipe Down the Docker 16 Storage and Handling 17 Disp...
Page 3: ... safety information or in service training contact your AMSINO sales representative or call AMSINO iReceptal Customer Service at XXXXXX Indications For Use The iReceptal Digital Surgical Suction System is intended to be used in the operating room pathology surgical centers and doctor s offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or ...
Page 4: ...ction System Customer Service Description The iReceptal Docking Station docker is a component of the iReceptal Digital Surgical Suction System The Stryker Rover rover another component of the system is a mobile unit used to suction and collect fluid waste and surgical smoke from a surgical site in an operating room After collection the rover is relocated and mated to the docker Once the rover is c...
Page 5: ...o not use iReceptal Digital Surgical Suction System in the presence of magnetic resonance imaging MRI devices Electrical Safety Use only AMSINO approved system components and accessories unless otherwise specified Using other electronic components and accessories may result in increased electromagnetic emissions or decreased electromagnetic immunity of the system Take special precautions regarding...
Page 6: ...a réglementation d Industrie Canada le présent émetteur radio peutfonctionner avec une antenne d un type et d un gain maximal ou inférieur approuvé pour l émetteur par Industrie Canada Dans le but de réduire les risques de brouillage radioélectrique à l intention des autres utilisateurs il faut choisir le type d antenne et son gain de sorte que la puissance isotrope rayonnée équivalente p i r e ne...
Page 7: ...id waste from the rover when the rover is connected to the docker H Water Inlet Port Allows fresh water to enter the rover when the rover is connected to the docker I Fixing Casters four Allows adjust the height of each docking station to align with the rover Increase caster height by turning the dial counterclockwise Decrease caster height by turning the dial clockwise J Power Cord Allows for the...
Page 8: ...For more information see Technical specifications Symbol Description ON POWER OFF POWER ALTERNATING CURRENT AC General warning Consult instructions for use Follow instructions for use This symbol is located near the protective ground locations on this device DETERGENTADD PORT WASTE OUTLET PORT WATER INLET PORT ...
Page 9: ...wer cord Configurations may vary See the Accessories section for power cord options NOTES Only individuals trained and experienced in the maintenance of reusable medical devices should install inspect and test this equipment The docker is installed in a utility closet or disposal area with access to electrical power a water supply and a fluid waste drain The healthcare facility is responsible for ...
Page 10: ... and against a wall with access to electrical power water and a fluid waste disposal drain b Recommended Install mounting screws not supplied through the mounting bracket of the docker and secure the docker to the wall NOTE If allowed by current local building and electrical codes mounting the docker to the wall is strongly recommended The docker is not to be installed in a patient environment c C...
Page 11: ...nnections for any leaks Repair any plumbing to stop leakage if necessary c Unscrew the lid of the detergent canister and pour the detergent into the detergent canister see the instructions for use supplied with the Docking Detergent REF iRC003 See the Accessories section d Mount the iReceptal 3 Docking Station Instruction Poster to a wall near the docking station Figure 8 To Apply Power Water and ...
Page 12: ...e rover to facility electrical power using the power cord b Push the power switch to the ON position c Read the WARNING message on the user interface then tap the OK key to access the CONTROL screen d Install a disposable manifold into manifold port e Attach a suction tube to one manifold port on manifold Make sure all the other manifold ports are capped f Place the attached suction tube into a si...
Page 13: ...er attach automatically b As the rover and docker attached together the Wash Cycle View will display automatically From the Wash Cycle View select a cycle by tapping the related key For the wash cycle options refer to the wash cycle options table Figure 11 To perform a wash cycle NOTES During the first rover docking procedure the information on the user interface display may appear inconsistent Th...
Page 14: ...he detergent with water 6 minutes Quick Drain Cycle drains the contents of both chambers 3minutes Extended Wash Cycle drains the contents applies detergent to the interior walls of both chambers and rinses the detergent with water Intermittent periods of soaking occur during the cycle 50 minutes c During the wash cycle inspect all the docker plumbing and connections for any leakage If leakage occu...
Page 15: ...pparent replace the equipment Check power cord for cuts Check power cord receptacle for bent pins or bent contacts Six months Check the plumbing connections and hoses for leaks Repair any plumbing to stop leakage as required Replace leaking hoses as required As required Check the two infrared communication ports for any obstructions Remove any obstruction covering the infrared communication ports ...
Page 16: ... minimum gown gloves face eye shield Soft lint free cloth Environmental Protection Agency EPA registered disinfectant with a claim for activity against Hepatitis B The following disinfectants have been validated for use with the AMSINO iReceptal 3 Waste Management System Sodium Hypochlorite Based Clorox Clean Up Disinfectant Cleaner with Bleach EPA Reg 67619 1 Quaternary Ammonium Based CaviCide EP...
Page 17: ...ically correct disposal to prevent dispersion of potential pollutants into the environment Please hand in the device at the end of its useful life to the local collection and recycling point for electrical and electronic devices Failure to comply may cause environmental contamination or infection and result in personal injury Per the European Community EC Waste Electrical and Electronic Equipment ...
Page 18: ...ctrical equipment to maximize the distance between the equipment Increase spatial distance Plug equipment into different outlets Error Messages NOTE For error message information see the instructions for use supplied with the rover Specifications Model REF iRD301 01 Electrical Power Requirements 100V 240V 50 60 Hz 3 A Power inlet module Power switch with 250V fuses on neutral and line connection E...
Page 19: ...ead MHT fitting and has a dedicated shutoff valve Potable tap water Approximately 34 liters 9 gallons per rinse cycle at default settings on standard cycle water usage fluctuates due to selected cycle and facility flow Refer to local plumbing codes to determine whether an external backflow prevention device is required Drainage Requirements Floor drain or permanent service connection per local plu...
Page 20: ...ake mitigation measures such as reorienting or relocating the iReceptal 3 Dockers REF iRD301 01 and REF iRD302 01 or shielding the location Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacturer s declaration electromagnetic immunity The iReceptal 3 Dockers REF iRD301 01 and REF iRD302 01 are intended for use in the electrom...
Page 21: ...itter Recommended separation distance d 1 2 P 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than...
Page 22: ...nd REF iRD302 01 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the iReceptal 3 Dockers REF iRD301 01 and REF iRD302 01 as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Rated maximum output power of transmitt...