![Amsino AMSure User Manual Download Page 77](http://html1.mh-extra.com/html/amsino/amsure/amsure_user-manual_2936262077.webp)
File
Name
IFU of
AMSure
®
Enteral
Feeding
Pump and Feeding Set
REF
#
L10196_US
Version
B
Page
77
of
81
AMSINO Confidential
14. Waste Disposal
14.1 Battery
Please follow local regulations to dispose of used batteries.
14.2 Enteral feeding set
After use, it shall be disposed of in accordance with relevant medical waste
disposal regulations.
14.3 Pump
The designed service life of the pump is 5 years. After the service life is
exceeded, it should be disposed of in accordance with relevant medical waste
disposal regulations.
15. Electromagnetic Compatibility
The
AMSure
®
Enteral Feeding Pump has been built and tested according to
60601-1-2:2020. FCC to be updated.
The pump is intended for use in the electromagnetic environment specified in
the table. The user of the pump should ensure that it is used in such an
environment.
Electromagnetic disturbances could cause disruption or malfunction of the
pump with essential performance alteration.
Basic performance
Item
Description
Internal power operation
Start normal operation at 25mL/h, the feeding accuracy
error is within ±10%, and the operation is normal during the
process, there should be no abnormal phenomena and
malfunctions
Grid power connection
operation
Start normal operation at 25mL/h, the feeding accuracy
error is within ±10%, and the operation is normal during the
process, there should be no abnormal phenomena and
malfunctions
Guidelines and manufacturer's declaration-electromagnetic emissions
Guidelines and manufacturer's declaration-electromagnetic emissions
The equipment is expected to be used in the following electromagnetic environment, and
the purchaser or user should ensure that it is used in this electromagnetic environment:
Test
Compliance
Electromagnetic environment-guide
Radio frequency emission
CISPR 11
1
The device uses RF energy only for its
internal functions, so its RF emission is
very low, and the possibility of causing
interference to nearby electronic devices is
very small
Radio frequency emission
CISPR 11
Class A
The equipment is suitable for use in non-
domestic and all facilities that are not