File
Name
IFU of
AMSure
®
Enteral
Feeding
Pump and Feeding Set
REF
#
L10196_US
Version
B
Page
74
of
81
AMSINO Confidential
Amsino shall not be liable for loss of or damage to the device during transport
to our customer service center. With the exception of the obligations set out in
this limited warranty, Amsino shall not be liable for direct or indirect damage or
damage by natural forces, regardless of the damage, whether based on an
agreement or the law or any other legal theory, even if Amsino has been
informed of the potential damage.
If, upon receipt, the pump packaging is damaged, user must ensure it is noted
on the courier’s paperwork.
13. Characteristics of ALARM SIGNALS
Alarm light indicator characteristics
Alarm type
Indicator Color
Flashing Frequency
Duty Cycle
High priority alarm signal
Red blinked
2Hz
50%
Medium priority alarm signal
Yellow blinked
2Hz
50%
Low priority alarm signal
Solid Yellow
Normally open
100%
Volume of auditory alarm signals
Unit
:
dB(A)
Alarm
Condition
Volume level
setting
Positions as
specified in
Table F.1 of
ISO 3744
Measured
sound
pressure
level
(dB)
A-weighted
sound
pressure
level
averaged
over the
measurement
surface (dB)
Remarks
HIGH
PRIORITY
Maximum
2
77.9
78.6
4
78.0
6
76.6
8
78.0
10
80.2
12
80.6
Minimum
2
76.4
75.1
4
73.4
6
73.3
8
73.9
10
77.1
12
76.6
HIGH
PRIORITY
Maximum
2
80.5
75.2
4
71.9
6
75.0
8
72.1
10
75.7
12
75.9
Minimum
2
70.9
68.1
4
65.7
6
65.8