If the wrong feeding set (with RFID tag) or damaged RFID tag is used, the
system will alert
“Incorrect Tubing” on status bar.
Fig 7.1.4-5 Incorrect Tubing
Caution
Feeding set identification system is an optional function on the pump. User can
switch on/off it in system maintenance page, referring to 7.3.5.2 RFID Switch.
The default setting is switch on.
RFID tag can record the feeding set usage time. When set usage>24 hours:
1) The prompt that the tubing used > 24 hrs. will always be rolling display on
the status bar to remind user to replace a new feeding set.
2) It will not interrupt the feeding task until feeding complete alarm occurred.
3) If new feeding button is touched, it will end current feeding task. The pump
cannot be used for feeding until a new feeding set is replaced to clear this
prompt. If user press start/stop button, a prompt will pop up to remind
“Please swipe RFID tag”
4) When alarm occurs, the alarm will come out first instead of the RFID
prompt info. The RFID prompt will pop out once alarm is eliminated.
7.1.5 Load Feeding Set
1.
Close the roller clamp and insert the drip chamber of the feeding set on
the drip chamber socket.
2.
Roll and stretch the silicone tube around the rotor, seat the adaptor in
socket.
3.
Insert downstream tube into slot.
The final assembly is as below: