6-26
129373-496
Operating Instructions
Sterilizer Operation
• Complete
– The complete tone sounds and the cycle summary and end-
of-cycle messages are printed. Display returns to the main menu (screen
#1).
6. Once the sterilizer completes and passes the Leak Test cycle, the unit can
be safely used for weekly processing.
NOTE: Sequential records of tests should be kept to detect if any major changes
in leak rates are occurring. Maintenance can then be scheduled to correct any
loose fittings, bad gaskets, etc.
6.7 Dart
®
Warm-up
Cycle Operation
The Dart Warm-up cycle is designed to bring the chamber up to operating
temperature in preparation for the DART (Bowie-Dick) Test cycle.
Dart Warm-up cycle is preprogrammed, and cycle parameters are fixed and
cannot be changed by the customer.
The Dart Warm-up cycle should be run as the first cycle of the day, prior to
performing a DART Test cycle.
1. Before running this cycle, refer to "Before Operating Sterilizer" at beginning
of this section.
NOTE: If sterilizer is equipped with double doors, the interlock type for Dart
Warm-up cycle is factory set and fixed at Type #1. Refer to S
ECTION
8 for
description of interlock types.
2. Press
CYCLE SELECT
on screen #1.
Pressing OPTIONS advances display to
the first Out-of-cycle Options menu
(screen #13).