12-6
129373-496
Operator Manual
Troubleshooting
Occurs if chamber
temperature exceeds
the maximum sterilize
temperature (Control
temp. + over temp.
value).
Screen with Service Instructions
12.2.6 Over Sterilize
Temperature
STATUS..
SERVICE INFORMATION:
236
OVER STERILIZE TEMPERATURE
→
→
→
→
→
STERILIZE TEMPERATURE IS ABOVE
SETPOINT BY MORE THAN PRESCRIBED
AMOUNT
CAUSES AND CORRECTION:
1.
STEAM PRESSURE MORE THAN 50 psig
→
→
→
→
→
CHECK STEAM SUPPLY PIPING
2.
CHAMBER STEAM TRAP MALFUNCTION
→
→
→
→
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S09
→
→
→
→
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
SERVICE
STATUS
PAPER
EXIT
MODE
FEED
Alarm
Description
Screen with Operator Instructions
NOTE: The control temperature is
the value the sterilizer maintains
during the sterilize phase. This
value is 2°C above the cycle steril-
ize temperature set point. The over
temperature value is added to the
control temperature value to arrive
at the temperature which causes
this cycle alarm.
SILENCE
SERVICE
ALARM
HELP
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
STATUS..
ALARM!
235
OVER STERILIZE TEMPERATURE
ABORT