Operator Manual
Introduction
(Continued on following page.)
Factory-Set Cycles and Cycle Values
1
The AMSCO C Series 16C and 20C
Prevacuum
Small Steam Sterilizer 16 x 16" (406 x 406 mm) and 20 x 20" (508
x 508 mm) is equipped with the following factory programmed sterilization cycles and cycle values (
Table 1
).
NOTE:
Default cycles for SFPP sterilizer shown in Table 1B on following page.
1. Supervisor must enable access to optional cycles, such as the Liquids Cycle.
Table 1. Factory Default Cycles, Prevacuum Configuration
Cycles:
Sterilize
Temp.
Sterilize
Time
Dry
Time
Recommended Load
*
Validation
Standard
1. Immediate Use
Prevac Cycle
270°F
(132°C)
4 Minutes
1 Minute
Unwrapped instrument tray containing
a single instrument; or multiple
unwrapped instrument trays, porous
and non-porous loads. Maximum
weight per tray: 25 lb (11.3 kg)
ST-8:2008
2. Prevac Cycle
270°F
(132°C)
4 Minutes
30 Minutes Double-wrapped instrument trays
containing multiple instruments,
porous and non-porous loads.
Maximum weight per tray: 25 lb
(11.3 kg)
ST-8:2008
3. Prevac Cycle
275°F
(135°C)
3 Minutes
30 Minutes Double-wrapped instrument trays
containing multiple instruments,
porous and non-porous loads.
Maximum weight per tray: 25 lb
(11.3 kg)
ST-8:2008
4. Prevac Cycle
270°F
(132°C)
4 Minutes
20
Minutes
**
Fabric Packs.
ST-8:2008
5. Immediate Use
Gravity Cycle
270°F
(132°C)
10 Minutes
1 Minute
Unwrapped instrument tray containing
a single instrument; or multiple
unwrapped instrument trays,
non-porous load. Maximum weight per
tray: 25 lb (11.3 kg).
ST-8:2008
6. Immediate Use
Gravity Cycle
270°F
(132°C)
3 Minutes
1 Minute
ST-8:2008
*
See
for recommended quantities.
** Dry time can be reduced to 5 minutes for a single fabric pack.