![Amsco 36 H Operator'S Manual Download Page 57](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188057.webp)
6-9
Sterilizer Operation
Operator Manual
P129394-068
6.9 Prevac Cycles
Prevac cycles,
270°F (132°C),
are used for sterilizing double-
wrapped instrument trays or fabric packs. Prevac cycles, 275°F
(135°C), are used for sterilizing double-wrapped instrument trays
only.
1.
Before running this cycle refer to
.
2.
See instructions for using the loading car/transfer carriage or
rack and shelves, earlier in this section, for procedures to load
the sterilizer.
•
For 270°F (132°C) Cycle, press 270 PREVAC button to start
the prevacuum cycle. Refer to
for descriptions of
cycle use and loading recommendations.
•
For 275°F (135°C) Cycle press 275 PREVAC touch-screen
pad to start the 275°F (135°C) Prevacuum cycle. Refer to
for descriptions of cycle use and loading
recommendations.
NOTE:
If the wrong cycle has been selected, see
3.
Sterilizer automatically progresses through cycle, as follows:
NOTE:
The control display indicates cycle progress in two ways:
a.
Large numerals in the screen’s center estimate remaining
cycle time,
b.
A change of phase name in the top status line of the screen.
ACTIVATE SEAL – Steam enters the door seal, pressing it
against inside surface of the door.
PURGE – Chamber is purged with steam. Start of Condition is
printed.
VACUUM/PRESSURE PULSES – Differing by temperature rating
for the cycle:
•
270°F (132°C) #1 - #4 – Four vacuum pulses.
•
275°F (135°C) #1 – #3 – Three vacuum pulses.
CHARGE – Chamber is charged with steam. Start of steam
charge is printed.
STERILIZE – Start of sterilize exposure is printed when the
chamber reaches sterilization temperature. Chamber
temperature is printed every minute.
FAST EXHAUST – Start of exhaust is printed and chamber is
exhausted to 4 psig (0.28 bar).
DRY – Start of dry is printed and display counts down dry time
remaining. Minimum prevacuum dry time defaults to one minute
to remove residual vapor at cycle end.
AIR BREAK – Chamber is returned to atmospheric pressure.
RETRACT SEAL – A vacuum is drawn on the seal, retracting it
from inner surface of door.
COMPLETE – Complete tone sounds. Cycle summary and end of
cycle messages are printed.
4.
Unload sterilizer (see instructions for using the loading car/
transfer carriage or rack and shelves,
for procedures to remove load).
WARNING – BURN HAZARD:
•
Sterilizer, rack/shelves and
loading equipment will be hot
after cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
•
Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door is
opened to minimize contact
with steam vapor.
WARNING – SLIPPING
HAZARD: To reduce slippery
conditions, keep floors dry by
immediately wiping up any
spilled liquids or
condensation in sterilizer
loading or unloading area.
WARNING – PERSONAL
INJURY and/or EQUIPMENT
DAMAGE HAZARD: When
closing the chamber door,
keep hands and arms out of
the door opening and make
sure opening is clear of
obstructions.
!
!
!
Summary of Contents for 36 H
Page 2: ......
Page 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Page 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Page 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Page 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...