Depending on circumstances the operator may choose to retry the single frequency, or cancel the single
frequency test.
Check that the patient fully understands the instructions previously given regarding use of the response
switch. Check also that the response switch is connected and operating and that tones are being output from
the headphones.
7.4.
RESPONSE SWITCH HELD ON
If the audiometer detects that the response switch is not released by the patient during an automatic test the
tone level will decrease to minimum and then the test will pause with the following message displayed on the
PC screen:
Depending on circumstances the operator may choose to abort the test, retry the frequency, or ignore the
error (and proceed to testing of the next frequency).
For a single frequency test, the tone level will decrease to minimum and then the test will pause with the
following message displayed on the PC screen:
Depending on circumstances the operator may choose to retry the single frequency, or cancel the single
frequency test.
Check that the patient fully understands the instructions previously given regarding use of the response
switch. Check also that the response switch is connected and operating and that tones are being output from
the headphones.