Chapter 2 System Overview
Part Number: 0160-6411, Rev A, 02/2019
Page 39 of 314
Software Release: cOS 5.00
trained
OPTIMEDICA
personnel will re-install and test the system to ensure that it is operating
within specification.
2.8 System Basics
Starting Up the System
1. Verify that the system components are properly connected and that the main power
cable is connected to an appropriate wall power outlet, as described in
.
2. Ensure that the system main power circuit breakers are in the ON position.
3. Insert the key into the key switch. Turn the key clockwise to the
(START) position and
immediately release. It may take up to 2 minutes for the computer screen to appear
active.
WARNING
Do not hold the key in the start position for more than two seconds. If the key is held in the start
position for more than two seconds, the system will do a hard shutdown. In the event of a hard
shutdown, you must contact the
OPTIMEDICA
Service Department at +1.800.511.0911.
Restarting the system will require up to a 45‐minute warm‐up period.
A system self-test and start-up routine begins, and the following screen displays on the system
control panel:
Figure 2.9 Start-up Screen
When the system self-test and start-up is complete, the Login Screen displays.
WARNING
If the control panel remains blank for an extended period of time during system start‐up, press
the Power button on the front of the control panel to turn on the control panel. If the control
panel remains blank, turn off the system with the key, wait at least one minute, and then restart
the system. If the screen is still blank, turn off the system and contact
OPTIMEDICA
Service.