Chapter 2 System Overview
Page 34 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
*
OPTIMEDICA
reserves the right to make changes to the device(s) herein. Device(s), therefore, may not
agree in detail with the published design or specifications. All specifications are subject to change with‐
out notice. Contact
OPTIMEDICA
or your local
OPTIMEDICA
representative for information on changes
and new products.
2.7 System Installation and Setup
Overview
The
CATALYS®
System is intended for use by trained Ophthalmologists for the treatment of
cataracts. The system is designed for installation and use inside a sterile operating room or in a
non-sterile pre-op area within an ambulatory surgery center (ASC) or a hospital. System
installation and testing will be performed by
OPTIMEDICA
-certified personnel.
The
CATALYS®
System is equipped with a detachable 3-wire AC power cord. When selecting the
location for system installation, a correctly grounded power connection is required for safe
system operation. Therefore, it is mandatory that the AC wall power outlet is correctly grounded
and is rated for the electrical load of the system. Turn off the system circuit breakers and
remove the system electrical power cord to isolate the system from mains power source. Follow
local electrical codes to ensure proper grounding of the AC wall power outlet. An appropriate
location for the
CATALYS®
System must be selected such that it can accommodate the system
size and allow for easy access by the patient, physician, nurse and/or assistant, and
OPTIMEDICA
Service Personnel. The rear of the system must remain clear after installation to provide access
to the system power cord, circuit breakers, and remote interlock connections.
Proper system operation requires the selection of a well-ventilated space in an office or surgery
room with a temperature-controlled environment. The system will function at ambient
temperatures between 15 °C (59 °F) and 32 °C (89.6 °F), with controlled relative humidity
between 0% and 80% non-condensing. The Heating Ventilation and Air Conditioning (HVAC)
system used to control the temperature in the installation environment must be able to control
the temperature within ± 2 °C (±3.6 °F) of the temperature set-point for the system to operate
correctly. If the temperature of the installation environment changes by more than ± 2 °C
Environmental Requirements (Non-operating)
Maximum altitude
Standard commercial shipping altitude
Temperature range
14 – 131°F (-10 – 55°C)
Maximum humidity
Up to 80% @ 131°F (55°C) non-condensing
Physical Characteristics
Height
1.15 m ± 0.025 m
Length
1.64 m ± 0.025 m
Width
0.87 m ± 0.025 m
Weight
340 kg (750 lbs)
Optical NOHD
0 m
Laser safety eyewear
Not required
IO
Display
GUI: touchscreen, flat panel
Power cable length
≥ 8.2 ft (2.5 m), detachable