Chapter 2 System Overview
Part Number: 0160-6411, Rev A, 02/2019
Page 25 of 314
Software Release: cOS 5.00
LIQUID OPTICS™ Interface Circular Label Reader
The
LIQUID OPTICS™
Interface circular label reader reads data from the circular label on the
LIQUID OPTICS™
Interface packaging and transfers it to the
CATALYS®
System software for
identification and tracking purposes. The circular label must be held flat against the label reader
on the left side of the system. Once the label has been read by the system, one treatment will be
enabled on the
CATALYS®
System. Laser treatment cannot be initiated until a new circular label
has been scanned.
Mobile Patient Bed or Patient Chair
The
CATALYS®
System can be used with either the Mobile Patient Bed or patient chair. Both the
Mobile Patient Bed and patient chair incorporate an articulating headrest, backrest, seat, and
footrest that are individually adjustable, as well as a precision movement joystick that adjusts
the bed position in three axes (x, y, and z).
Additionally, the Mobile Patient Bed can be wirelessly connected to the
CATALYS®
System via
Bluetooth
1
to position a patient for Laser Cataract Surgery. Once the laser procedure is
completed, the Mobile Patient Bed can be rotated out from under the
CATALYS®
System to
perform surgical removal of the crystalline lens and IOL installation, or the bed can be used to
transport the patient to a surgery suite. The Mobile Patient Bed is equipped with an on-board
battery that is charged from an external AC power source.
1. The Mobile Patient Bed and
CATALYS
System are equipped with a class 2 Bluetooth radio.