Chapter 4 CATALYS® System Procedure
Page 204 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Removing the Suction Ring and Disposable Lens
After the system detects that the Mobile Patient Bed has been disconnected from the
CATALYS®
System or the patient chair has been unlatched, the Capture panel opens. As instructed in the
Capture and Disposable Lens panels of the Undocking Screen:
1. Press the
(Capture)
O
(off) button on the docking keypad (located on the system
front panel directly above the patient) to release the suction ring.
Figure 4.47 Turn Off Capture
2. Remove the suction ring and the disposable lens from the system, and dispose in
compliance with local regulations and applicable municipal codes.
Capture OFF
Button