Chapter 4 CATALYS® System Procedure
Part Number: 0160-6411, Rev A, 02/2019
Page 177 of 314
Software Release: cOS 5.00
Installing the Disposable Lens
1. Open the
LIQUID OPTICS™
Interface packaging by peeling back the Tyvek® package lid.
Figure 4.17
LIQUID OPTICS™
Interface
2. Remove the disposable lens from the packaging.
3. While looking at the underside of the disposable lens carriage, bring the disposable lens
up to the disposable lens carriage.
4. With
gentle
upward pressure, twist the disposable lens in the direction of the Mobile
Patient Bed or patient chair until the disposable lens tabs line up into the disposable lens
carriage. The disposable lens should seat flush in the groove.
Figure 4.18 Installing the Disposable Lens (steps 3 and 4)
5. Turn the disposable lens in the opposite direction (towards the touchscreen, or away
from the Mobile Patient Bed or patient chair) until you feel a stop, and at the same time
gently
pull the disposable lens towards the floor.
Fluid Reservoir
Suction Ring
Disposable Lens
Steps 3 and 4