Chapter 3 Software Navigation
Page 122 of 314
Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Vacuum Panel
After the system verifies installation of the disposable lens, the Vacuum panel opens. As
instructed in the panel, place and center the suction ring of the
LIQUID OPTICS™
Interface on
the patient’s eye, and then apply patient vacuum.
NOTE
Refer to
for detailed instructions on placing the suction
ring and applying patient vacuum.
Figure 3.76 Docking Screen with Vacuum Panel Open
When the system detects that the suction ring has been placed and patient vacuum applied, the
system emits an audible “ding” and a check mark displays in the Vacuum panel. As instructed in
the Vacuum panel, add sterile buffered saline solution to the suction ring.