Specifications
| 2-3
Sample System Limits
Sample Pressure
Up to 10,000 kPag (1,450 psig / 100 barg) can be accommodated, depend-
ing on the application. See product markings or the Final “As-Built” draw-
ings in the analyzer Documentation Package.
Oven/Sample System Enclosure Temperature
For applications with a heated sample system, the Oven will operate at
temperatures up to 150 °C (302 °F).
Sample Transport
Gas Applications:
By aspiration, using Instrument Air or N
2
as the drive gas.
Liquid Applications:
Using sample stream pressure.
Minimum 5 psi differential pressure inlet to outlet.
Instrument Air Requirements
The aspirator medium must be transported in 316 stainless steel.
Minimum pressure:
490–700 kPag (70–100 psig / 4.8–6.9 barg)
Air Quality:
As per ANSI/ISA-S7.0.01 (1996) Quality Standard for Instrument Air.
Nitrogen Requirements:
If using N
2
, use only ultra high purity (UHP) Nitrogen.
For “application specific” N
2
requirements, refer to the Final “As-Built”
drawings in the analyzer Documentation Package.
Sample Fluid Flow Rate
Application specific.
Summary of Contents for Western Research IPS-4
Page 48: ...2 18 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 140: ...4 48 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 172: ...5 32 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 178: ...6 6 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...
Page 218: ...A 2 IPS 4 Dual Bench UV IR Analyzer Analyzer Light Path UV Optical Bench Schematic...
Page 219: ...Appendix A Drawings A 3 Analyzer Light Path NDIR Optical Bench Schematic CONDENSING...
Page 222: ...A 6 IPS 4 Dual Bench UV IR Analyzer Detector Board 100 2046...
Page 223: ...Appendix A Drawings A 7 Display Interface Board 100 2049...
Page 225: ...Appendix A Drawings A 9 AC Disconnection Board 100 2077...
Page 227: ...Appendix A Drawings A 11 Relay Board 100 2050...
Page 228: ...A 12 IPS 4 Dual Bench UV IR Analyzer Analog Board 100 2047...
Page 229: ...Appendix A Drawings A 13 MCU Board 100 2045...
Page 230: ...A 14 IPS 4 Dual Bench UV IR Analyzer Xenon Lamp Power Supply Board 100 2061 UV Optical Bench...
Page 231: ...Appendix A Drawings A 15 Infrared Source Assembly Interface Board 100 2838 IR Optical Bench...
Page 232: ...A 16 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 1 of 5...
Page 233: ...Appendix A Drawings A 17 Electronics Enclosure Wiring Diagram Sheet 2 of 5...
Page 234: ...A 18 IPS 4 Dual Bench UV IR Analyzer Electronics Enclosure Wiring Diagram Sheet 3 of 5...
Page 235: ...Appendix A Drawings A 19 Disconnect Enclosure Wiring Diagram Sheet 4 of 5...
Page 236: ...A 20 IPS 4 Dual Bench UV IR Analyzer Disconnect Enclosure Wiring Diagram Sheet 5 of 5...
Page 238: ...S 2 IPS 4 Dual Bench UV IR Analyzer This page intentionally left blank...