Dansensor
®
MAP Mix Provectus
Relay signalling
The following describes how relays act during power OFF, normal and fault states:
I/O signals for machine control
I/O machine control:
Start/Stop input
Fault relay
Alarm relay
FAULT Relay
State of device
Device OFF
DSUB pin 11 connected to pin 12
“Fault” (Off )
Device ON - OK
DSUB pin 11 connected to pin 13
“OK”
Device ON - Fault
DSUB pin 11 connected to pin 12
“Fault”
DSUB pins/colour
Pin 11 - Grey/Pink (Common)
Pin 12 - Blue/Red
Pin 13 - White/Green
ALARM Relay
State of device
Device OFF
DSUB pin 5 connected to pin 6
“Alarm” (Off )
Device ON - OK
DSUB pin 5 connected to pin 7
“OK”
Device ON - Fault
DSUB pin 5 connected to pin 6
“Alarm”
DSUB pins/colour
Pin 5 - Grey (Common)
Pin 6 - Pink
Pin 7 - Blue
Simplified schematic of control inputs
Simplified schematic of relay outputs
Isolation barrier
Isolation barrier
Bipolar input
10-32 VDC
Bipolar input
10-32 VDC
Relay output
48 VDC
1A max.
Relay output
48 VDC
1A max.
Normally
Closed (NC)
Normally
Closed (NC)
Common
Normally
Open (NO)
Normally
Open (NO)
+/-
-/+
Summary of Contents for mocon Dansensor MAP Mix ProVectus
Page 1: ...P N 300862 K 11 2019 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 28: ...26 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 72: ...70 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 79: ...This blank page has been inserted to enable double sided printing of the document...