Dansensor
®
LeakPointer
H
2
O
EN
User Guide
11
Compressed air
NOTE! The compressed air supply hose is not included.
Connect the compressed air supply hose
to the inlet
at the back of the unit.
Make sure
- that the hose is pushed into the fitting as far as it goes.
- to check the connection by pulling the hose slightly outwards.
- that the hose and the compressed air complies with the specifications in
CAUTION! When connecting the hose, make sure that it can not be squeezed or
blocked in any way.
CAUTION! If the specified maximum pressure is exceeded, there is a danger of
bursting of tubing and other components.
NOTE! If the supplied air pressure is outside the specified range, the ejector
efficiency will be reduced.
Pressure regulator/filter
We recommend to use a pressure regulator/filter unit which ensures that the air quality fulfills
the required specifications.
NOTE! Make sure that the hose between the device and the pressure regulator is
max. 1m, otherwise you must compensate for the pressure drop in the hose.
NOTE! A regulator/filter unit does not ensure that the air is dry and therefore it is
recommended to install a type of air-drying unit as well.