![Ametek Mocon Dansensor LeakPointer 3 User Manual Download Page 9](http://html1.mh-extra.com/html/ametek/mocon-dansensor-leakpointer-3/mocon-dansensor-leakpointer-3_user-manual_2931374009.webp)
Dansensor
®
LeakPointer 3/3
+
EN
User Guide
7
1. Introduction
General description
Dansensor® LeakPointer 3
and
Dansensor® LeakPointer 3
+
are table top off-line leak
testers for Modified Atmosphere Packages (MAP).
The devices can fast and easily discover seal integrity problems on most types of modified
atmosphere packages with a minimum of 10% CO
2
in the gas mix.
The devices operate with a very simple measuring principle:
A CO
2
filled package is placed in a chamber in which a vacuum is built up. If there is a leak in
the package, the gas from the package will leak into the chamber, thus the increase in the CO
2
content can be measured.
The PackFix mechanism (available either as standard built-in or as a retrofitted option)
enables the devices to detect a leak and to calculate the leak size (according to the leak size
definitions of the Dansensor® Leak Simulation Kit).
All measurement data can be saved in the device’s database and exported for further analysis.
Summary of Contents for Mocon Dansensor LeakPointer 3
Page 1: ...P N 380064 F 06 2021 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 8: ...6 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 20: ...18 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 94: ...92 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 98: ...96 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 99: ...This blank page has been inserted to enable double sided printing of the document...
Page 100: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...