Setting up products for “Multi Package” testing
The setup procedure for
Multi Package
products includes the following:
Create product - see
Determination of vacuum setting
1. Prepare the test package for DeltaP measuring as described in
.
NOTE! Make sure that neither the needle nor the hose is hit by the chamber
when the chamber is closed. We recommend to place the needle through the
side of the package.
2. Select
Workflow ->
Products ->
Edit
and then select the product you have
created for the specific package type.
3. Set
Mode
to
Multi Package
.
4. Tap the yellow
icon to go to the
Create Test Vacuum
screen.
5. Set the
Chamber vacuum (delta)
to
300 mbar
, then close the chamber to start the
measurement.
6. The chamber will be evacuated to the set value and the measured values are continuously
displayed on the screen.
NOTE! If the currently measured “CO2” value increases when vacuum is applied,
then the package is most likely leaking and you should thus use another
package to perform the test.
Summary of Contents for Mocon Dansensor LeakPointer 3
Page 1: ...P N 380064 F 06 2021 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 8: ...6 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 20: ...18 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 94: ...92 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 98: ...96 User Guide EN Dansensor LeakPointer 3 3 P N 380064 F 06 2021...
Page 99: ...This blank page has been inserted to enable double sided printing of the document...
Page 100: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...