The setup menu
SET.
1. Press
when
SET.
appears on the display.
2. Use
and
to set required value.
3. Press and hold for 3 seconds to change the value. (for typing in values, see
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
•
FLO.L
100
Flow alarm limit. Alarm limit for min. flow in ml/min.
•
FLO.H
150
Flow alarm limit. Alarm limit for max. flow in ml/min.
•
DEL.
3
Gas flushing time in seconds before activation of alarms.
Valid area 0 to 250.
•
CU.0
- - - -
O
2
value which must correspond to 0mA (0 V)
•
CU.4
0%
O
2
value which must correspond to 4mA (2 V)
•
CU.20
10%
O
2
value which must correspond to 20mA/10V
•
PA.
1001
Setting up of options, relay function and external measurement
control.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
When typing in values for
FLO.L
and
FLO.H
, error
ER.1
appears, if
FLO.L
is set to a higher value
than
FLO.H
.
Only one of the parameters
CU.0
and
CU.4
can be set, the other one is displayed as “- - - -”.
If identical values are entered for the 0 and 20mA or the 4 and 20 mA values, error code
er. 1
will be displayed.
Diagnostic menu
DIA.
1. Press
when
DIA.
appears in the display.
2. Use
and
to set required value. For a permanent readout of the chosen value press
and hold for about 3 seconds. The value wil now be displayed until you press ,
or
again.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
•
EMF.
165.0
Present sensor EMF in mV.
•
TPT.1
33
Present internal ISM-3 temperature in °C.
•
TPT.2
0
Present difference in heater temperature in °C. Difference between
present temperature and ideal temperature.
•
OFFS
-0.12
Sensor offset. Result of latest off set adjustment in mV.
•
TI.PO
1234
Hour counter in number of 10 hours. Number of hours in which ISM-3
has been switched on.
•
TI.CA
678
Hour counter in numbers of 10 hours. Number of hours since the last
calibration of ISM-3.
•
TI.FL
983
Hour counter in number of 10 hours. Number of hours in which the
flow has passed through the sensor (flow >30ml/min).
•
PRG.
1.0
Installed program version.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Calibration menu
CAL.
See the chapter about calibration.
Summary of Contents for mocon Dansensor ISM-3
Page 1: ...P N 200114 I 09 2021 UserManual...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 6: ...4 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 10: ...8 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 12: ...10 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 20: ...18 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 28: ...26 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 32: ...30 User Manual EN Dansensor ISM 3 P N 200114 I 09 2021...
Page 33: ...This blank page has been inserted to enable double sided printing of the document...
Page 34: ...0 0 VSPQF 4 3 OOFEFWFK 3JOHTUFE FONBSL 5FM JOGP EL NPDPO BNFUFL DPN XXX BNFUFLNPDPO DPN...