Maintenance and Troubleshooting
|
6-9
See Chapter 7 for
spare part ordering numbers.
Measuring Cell Maintenance
This procedure discusses a Standard Range Measuring Cell (40 cm or
smaller, Figure 6-3.1) and a Low Range/White Cell (81.2 cm, Figure 6-3.2).
See also Figure 6-2 to locate and identify parts. Other types of Measuring
Cells are discussed in
Manual Supplements
shipped with the analyzer.
Replace/clean parts in the Measuring Cell:
• Every year.
• Any time an unscheduled cleaning is performed.
• If the analyzer responds slowly to a Zero when the sample system is
free of restrictions.
• If evidence of contamination is present in the sample tubing.
“P/N” refers to Part Number.
To clean and replace parts in the Measuring Cell (see Figures 6-1, 6-2, and
6-3.1 or 6-3.2):
Hazardous Locations
Before proceeding, test the area around the analyzer for explosive
gases and proceed only when the area is found to be safe.
Do not remove the cover of the Explosion-Proof Digital Communica-
tions Port on the front of the Lower Enclosure, do not connect a serial
cable to the Communications Port, do not open the Upper/Lower En-
closures, and do not power up/down the analyzer or computer if there
is an explosive gas atmosphere present.
1. Connect a serial cable between the computer and the analyzer’s Com-
munications Port. Using the Configurator Software, change the
Flow
Control
setting to ‘
3
’ (
zero
). Click
OK
then
Apply
. Do not save the
changes to EEPROM.
2.
Analyzer with Heated Cells:
Remove the Insulating Jacket or open the Oven door to allow the Cell
to cool down enough to work on it.
Setup (tab)
Gas Calibration<<Flow
Control>>
NOTE
!
WARNING
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...