5-2 | Model 933S UV Analyzer
Calibrating the Analyzer and Gas Sensors
To calibrate the analyzer and gas sensor(s), you must be communicating
with the analyzer using the S93XS System 200 Configurator Software.
When calibrating the analyzer, you will use the
Gas Calibration
and
Ana-
lyzer Data
tab to set up and calibrate the analyzer. When calibrating the
sensor(s), you will use the
Gas Calibration
and
Sensors
dialog box to set
up and calibrate the sensor(s).
• The
Gas Calibration
dialog box (Figure 5-1) contains UV gas species
information and sample system control parameters. For more infor-
mation on setting up parameters, see the section called “Setting Gas
Calibration Parameters” in Chapter 4.
• The
Analyzer Data
tab (Figure 5-2) displays the Span gas readings.
• The
Sensors
dialog box (Figure 5-3) contains the sensor calibration
parameters, which will be factory-set for the sensor(s) used in your ap-
plication, on the corresponding screen (
Sensor 1
,
Sensor 2
, or
Sensor 3
– see “Sensors Setup” in Chapter 4 for the different Sensor screens).
The calibration procedure takes approximately 15–30 minutes to com-
plete. If configured, a concentration alarm may occur if the sensor fails
or goes significantly out of calibration.
The
Sensors
dialog box also allows you to enter the concentration
(in %) of the interfering gas(es) in the process stream. For example,
enter the amount of CO
2
(in %) in the
% in Process Gas Stream
field
for
Interference Gas #1
and enter the amount of H
2
O Vapour (in %)
in the
% in Process Gas Stream
field for
Interference Gas #2
.
Figure 5-3 shows the
Sensors
dialog box with
Differential
Signal
Type
selected on the
Sensor 1
screen. Depending on how the sensor is
configured for your application, data for the sensor will be displayed
on either the
Sensor 1
,
Sensor 2
, or
Sensor 3
screen.
NOTE
Summary of Contents for 933S
Page 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Page 34: ...2 10 Model 933S UV Analyzer...
Page 35: ...Specifications 2 11...
Page 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Page 37: ...Specifications 2 13...
Page 38: ...2 14 Model 933S UV Analyzer...
Page 39: ...Specifications 2 15...
Page 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Page 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Page 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Page 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Page 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Page 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Page 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Page 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Page 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Page 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Page 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Page 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Page 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Page 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Page 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...