Section 2
Page 2 - 13
Fixed Thermal Imaging System
Installation Guide
Fig. 12 Water Connection Details
RUGGED HOUSING
WATER
IN
WATER
OUT
FLOW RA
TE 1L/Min
CONNECT END OF HOSES TO RUGGED HOUSING
WATER INLET & OUTLET CONNECTIONS AS SHOWN
SECURE WITH SUITABLE CLIP
2
1
WATER CONNECTION DETAILS
BARBED HOSE OD 9.5mm (0.375") or 804207 1/8 BSP to
9.5 (3/8") AS SUPPLIED.
IF REMOVED REQUIRES
FITTINGS WITH G1/8" (1/8"BSP) MALE THREAD.
ANY ALTERNATIVE FITTING MUST HAVE A 7mm BORE
NOTE:
FLOW CAN BE REVERSED
ALWAYS USE FLEXIBLE PIPING TO ALLOW
PAN AND TILT MOVEMENT.
NOTE
WATER pH LEVEL MUST
BE WITHIN 5-8pH.
WATER HOSE No
SUPPLIED BY
DETAILS
CUSTOMER
1
TO SUIT 9.5mm (3/8") OD BARBED HOSE FITTING
CUSTOMER
2
TO SUIT 9.5mm (3/8") OD BARBED HOSE FITTING