background image

1 Introduction

Pump P-910 is a single channel peristaltic
laboratory pump for use as a sample pump to
fill sample loops in an ÄKTA

design 

chromatography system. Pumping action is
provided by a roller cage driven by a stepper
motor controlled from a PC running 
UNICORN

control system, version 2.10 or

higher.

Features:

Flow rates up to 5 ml/min.

Accurate and reproducible flow rate.

Safety

WARNING! When using hazardous
chemicals, all suitable protective measures,
such as protective glasses, must be taken.

2 Installation

CAUTION! Before connecting Pump P–910
ensure the power is switched OFF at the pump
or for the complete system.

Unpacking

Unpack the instrument and check the items
against the packing list. Inspect the items for
obvious damage which may have occurred
during transportation.

Installing the pump

At delivery, the pump is ready to use, with
only the tubing to be installed.

1

Mount the pump vertically,  in the way it

is shown in the user documentation of
your ÄKTAdesign chromatography sys-
tem.

2

Connect the pump with 
two UniNet cables as a 
part of the UnitNet 2

chain.

Assembling pump tubing

1

Cut the pump tubing squarely to length of
140 mm.

2

Insert one connector

in each end of the
tubing.

3

The distance between the two connectors
must be 140 mm. This ensures firm 

positioning of the tubing.

Instructions

Pump P-910

56-3000-47

Edition AD

140 mm

UniNet 2 chain

56-3000-47_AD.qxd  1998-10-21 14:17  Sida 1 (1,1)

Summary of Contents for P-910

Page 1: ...complete system Unpacking Unpack the instrument and check the items against the packing list Inspect the items for obvious damage which may have occurred during transportation Installing the pump At...

Page 2: ...a method or manually via the functions available in UNICORN Setting the flow rate 1 Select menu System Control Manual Pump in UNICORN 2 Select the instruction SampleFlow in the list 3 Enter the requir...

Page 3: ...ee section 2 Cleaning in place Pump a cleaning or sanitizing agent through the pump The standard recommendation is to pump 1 M NaOH for 30 minutes and then wash out with buffer WARNING NaOH is injurio...

Page 4: ...the roller bearings As the cage rotates the rollers rotate in the opposite direction driven by drive wheels which engage the chassis ring The roller diameter and the gear ratio ensure that the roller...

Page 5: ...declaration of conformity is valid for the instrument when it is used in laboratory locations used in the same state as it was delivered from Amersham Pharmacia Biotech except for alterations describe...

Page 6: ...6 Pump P 910 Instructions 56 3000 47 Edition AD 56 3000 47_AD qxd 1998 10 21 14 17 Sida 6 1 1...

Page 7: ...7 Pump P 910 Instructions 56 3000 47 Edition AD 56 3000 47_AD qxd 1998 10 21 14 17 Sida 7 1 1...

Page 8: ...otech AB SE 751 84 Uppsala Sweden Trademarks KTA and UNICORN are trademarks of Amersham Pharmacia Biotech Limited or its subsidiaries Amersham is a trademark of Nycomed Amersham plc Pharmacia and Drop...

Reviews: