Test record
7
ÄKTAexplorer Installation Guide 18-1139-59 Edition AC
29
7 Test record
Date: ................................................................................
ÄKTAexplorer serial no.: ..............................................
7.1
Gradient test result
Gradient linear from ................... %B to ..................%B. (10 - 90%)
7.2
Step response test result
Step response table:
7.3
UV response test result
UV response table:
1
Programmed
Conc.%B
2
Value
read
3
Baseline cor-
rected value
4
Normalised
value
5
Allowed
interval
100
95
94 - 96
70
69 - 71
30
29 - 31
5
4 - 6
0
1
Wavelength
(nm)
2
Value read
100% B 0% B
3
Baseline
corrected
value
4
Absorbance
ratio
5
Allowed
interval
254
265/254
1.11 - 1.26
265
265/280
1.26 - 1.53
280
18113959AC_TK.book Page 29 Monday, June 23, 2003 4:45 PM