Initializing Pump P-960 in UNICORN
1
Make sure that the correct strategy is installed.
2
In
UNICORN Main menu
, select
Admininstration:System Setup
.
3
Select the system used and click
Edit
.
4
Click
Component
.
5
Select
Sample pump P-960
and click
OK
.
6
Click
OK
again and then
Close
to complete the initialization.
2.8
Removing air from the sample flow path
Note: Make sure that any air in the sample flow path is removed before
beginning to use Pump P-960. Air in the sample pump or in the
flow path might cause pressure pulsation and inaccurate flow.
The procedure for filling the sample flow path is decribed in section
5.2.2 Removing air using a purge tubing. You can also find this
information in ÄKTAdesign Optional Configurations User Manual.
2.9
Connecting an air sensor to Pump P-960 (optional)
Air sensor Air-912N (standard), Air-912, Air-925 or Air-925N can be
used at the sample inlet before the pump to prevent air from entering
the flow path.
The signal cable from Air sensor Air-912N (or Air-925N) is connected
directly to the air sensor socket underneath the pump. That minimizes
the time required to stop the pump when air is detected.
For detailed instructions on how to install the air sensor in your
ÄKTAdesign system, refer to the ÄKTAdesign Optional
Configurations User Manual.
2.9.1
Initializing the air sensor in UNICORN
1
In
UNICORN Main menu
, select
Admininstration:System Setup
.
2
Select the system used and click
Edit
.
3
Click
Component
.
4
Select
Airsensor P-960
and click
OK
.
5
Click
OK
again and then
Close
to complete the initialization.