2
Installation
10
Pump P-960
User Manual 18-1172-73 Edition AA
2.4
Connecting UniNet-2 cables
Pump P-960 is controlled from a PC running UNICORN version 4.12
or higher via a UniNet-2 cable. Power is also supplied through this
cable.
The UniNet-2 chain connects, in series, Pump P-900/Pump P-920 with
Pump P-960 and other modules.
In ÄKTAdesign chromatography systems that include Pump P-960 as a
standard component, the UniNet-2 cables are installed at delivery.
To connect the UniNet-2 cable:
1
Remove the termination plug from the last component in the
UniNet-2 chain in the system.
2
Connect the supplied UniNet-2 cable
between the last component and the
UniNet-2 socket (the leftmost socket)
underneath Pump P-960.
The pump is now connected at the end of
the UniNet-2 chain. The termination
plug is not required since the sample
pump has an internal termination.
CAUTION!
The mains power to the ÄKTAdesign chromatography
system must be switched OFF before connecting the unit to the
UniNet-2 link.
18117273aa.book Page 10 Thursday, July 3, 2003 5:13 PM