7.
FAULTS AND FAULT FINDING
ALARMS will stop machine operation. It is not possible to restart the machine
without addressing the fault.
Alarm
Description
Trigger
Review and Check
Corrective Action
ALARMS
RO1 Pump Over
Pressure Alarm
Pump
pressure
transducer
1) RO1 Membranes are blocked
2) Concentrate valve on RO1 closed
3) Pressure switch faulty
1) Check valve status
2) Perform chemical clean,
RO2 Pump Over
Pressure Alarm
(DPH only)
Pump
pressure
transducer
1) RO2 Membranes are blocked
2) Concentrate valve on RO2 closed
3) Pressure switch faulty
1) Check valve status
2) Perform chemical clean,
RO1 Pump Fault
Alarm
1) Check and inspect pump RO1
1) Contact AmeriWater
service
RO2 Pump Fault
Alarm (DPH
only)
1) Check and inspect pump RO2
1) Contact AmeriWater
service
Temperature
High Alarm
Loop Over
Temp Alarm
Set-point
Exceeded
1) Supply to loop over limit - check
setting in CALIBRATION MENU > ALARM
SETUP screen LOOP OVER TEMP ALARM
2) Softened Potable incoming water
supply too hot
1) Check incoming water
supply
Output Quality
Alarm
Output Quality
Alarm Set-
point
Exceeded
1) Rinse period exceeded max time RO
PARAMETERS SCREEN AUTO RINSE
2) Loop supply quality over range
3) Loop supply line cell error
1) Check line cell
2) Perform chemical clean
Tank Quality
Alarm
Tank Quality
Alarm Set-
point
Exceeded
1) Tank line cell out of range RO SUPPLY
screen TANK QUALITY settings
2) Tank supply quality over range RO
SUPPLY screen LOOP QUALITY settings
1) Drain tank refill and
perform extra rinse
2)Check line cell
Water Loss
Alarm
Base Tank
Float Switch
1) Tank empty - check water supply to
machine
2) Check tank ball valve operational
3) Check tank float switches
4) Check inlet solenoid valve
Over
Temperature
Fault Alarm
Tank
temperature
probe
1) Tank over temperature check
2) Check loop return temp
3) Check loop supply temp
Emergency Stop
Activated Alarm
E Stop
1) Reset E stop on Control Panel
2) reset E stop on touch screen
PLC Analogue
Card Error Alarm
PLC
Contact AmeriWater
Contact AmeriWater
PLC to PCB
Comms Error
Alarm
Contact AmeriWater
Contact AmeriWater
Output High
Pressure Alarm
Delivery High
Pressure
Switch
1) Check if valve in the loop is closed
2) Check Permeate M. Valve is open
3) Verify Permeate relief valve is working
1) Increase setting on
pressure switch using
adjustment screw
Summary of Contents for MediQA MDP Series
Page 70: ...69 98 0157 Rev K 10 FLOW DIAGRAMS SPH UNITS ...
Page 71: ...70 98 0157 Rev K DPH UNITS ...
Page 72: ...71 98 0157 Rev K 11 WIRING SCHEMATIC SPH UNITS ...
Page 73: ...72 98 0157 Rev K ...
Page 74: ...73 98 0157 Rev K ...
Page 75: ...74 98 0157 Rev K ...
Page 76: ...75 98 0157 Rev K ...
Page 77: ...76 98 0157 Rev K ...
Page 78: ...77 98 0157 Rev K ...
Page 79: ...78 98 0157 Rev K ...
Page 80: ...79 98 0157 Rev K DPH Units ...
Page 81: ...80 98 0157 Rev K ...
Page 82: ...81 98 0157 Rev K ...
Page 83: ...82 98 0157 Rev K ...
Page 84: ...83 98 0157 Rev K ...
Page 85: ...84 98 0157 Rev K ...
Page 86: ...85 98 0157 Rev K ...
Page 87: ...86 98 0157 Rev K ...