51
98-0157 Rev K
6.5.1
FLOW METERS button
The Flow Meter menu allows access to the calibration settings.
OFFSET and GAIN are calibration parameters. Offset adds or subtracts a
constant value to the signal. Gain multiplies the signal by a constant factor.
6.5.2
RO PARAMETERS button
In the RO parameters menu, the target percent recovery for the RO can be set.
The default is 65% for stage 1 and 85% for stage 2 (if the unit is a dual pass).
The run time clock can also be reset from this menu.
Figure 48
Figure 49