7
98-0156 Rev G
Bringing the system to a stop in the event of an emergency
To shut the unit down in the event of an emergency rotate the
yellow/red door isolator switch to the “OFF” position. Isolate the 3-
phase electrical supply to the unit at source. There will be no
requirement to isolate any of the hydraulic lines as all valves close
when de-energised.
Safety information for maintenance tasks
The owner of the unit must take care to ensure that only authorized
personnel who have been sufficiently informed for the task at hand
perform all maintenance and inspection tasks.
Before beginning tasks on the electrical equipment of the system, a
check must confirm that power has been disconnected. In addition, the
system must be secured to prevent it from being turned on again
unintentionally. The isolator switch is capable of accepting a lock-out
padlock when in its “OFF” position. The unit may contain water at
scalding temperatures; ensure it is drained down and safe to work on
before carrying out any maintenance tasks.
Disposing of system parts and operating materials
Refer to Section 10.1 for details.
Unauthorized conversion and manufacturing replacement
parts
Conversion or modification of the system is only permitted with the
approval of AmeriWater. Original replacement parts authorized /
supplied by the manufacturer enhance safety and ensure design
performance. The use of unauthorized parts may void the warranty on
the unit, impair its performance or compromise the safety of those
operating it.
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...