198
Rev. 5
Appendix
Model 430IP
Note
Allow about 90-seconds (from power-up) for the TCP/IP link between the
Model 430 and host computer to be established.
In order to access the Model 430IP using
TCP/IP
, either the Model 430
System Name
or
IP Address
must be known. The
System Name
should be
available from the Model 430IP configuration documentation. If the
IP
Address
is assigned statically, it should also be available from the Model
430IP documents.
If the IP Address is assigned dynamically, it will not be known because it
changes on each Model 430 power-up; in this case the
IPNAME
must be
used (the
IPNAME
is also known as the
System Name
1
). With serial port
communication established (refer to “RS-232 Configuration” on page 110),
query the unit for
IPNAME
as follows (refer to “SCPI Command
IPNAME?
The port will respond with the
IPNAME
(
System Name
).
Open a web browser on the host computer. In the address field, type
“http://” followed by the
System Name
, and press <
ENTER
>. For example,
with a
System Name
A10123_X-AX:
The following initial screen should appear:
1. Refer to “Net Settings Submenu” on page 72.
Figure A-12.
http:// - System Name Entry
Figure A-13.
Initial Screen for Browser Access of the Model 430IP
Summary of Contents for 05100PS-430-601
Page 2: ......
Page 10: ...x Rev 5 List of Figures ...
Page 12: ...xii Rev 5 List of Tables ...
Page 18: ...xviii Rev 5 Foreword Safety Summary ...
Page 30: ...12 Rev 5 Introduction Operating Characteristics ...
Page 42: ...24 Rev 3 Installation Power Up Procedure ...
Page 114: ...96 Rev 5 Operation Summary of Operational Limits ...
Page 156: ...138 Rev 5 Remote Interface Reference Error Messages ...
Page 168: ...150 Rev 5 Service Return Authorization ...
Page 226: ...208 Rev 5 Index ...